Veritas Reverse Total Shoulder Replacement recall
restor3d Inc. recalled Veritas Reverse Total Shoulder Replacement on November 11, 2026 because of (Retroactively reported) Pairing of shoulder implant stem sizes 1 and 2 with the largest glenosphere diameter (40 mm) creates an excessive mechanical loading configuration in which certain polyethylene components lack adequate retention strength, potentially resulting in dislocation, reduced range of motion, or revision surgery. It was sold nationwide.
Why it was recalled
(Retroactively reported) Pairing of shoulder implant stem sizes 1 and 2 with the largest glenosphere diameter (40 mm) creates an excessive mechanical loading configuration in which certain polyethylene components lack adequate retention strength, potentially resulting in dislocation, reduced range of motion, or revision surgery.
What's recalled
Veritas Reverse Total Shoulder Replacement, Model/Catalog Number: 3500-140C3, 3500-140U3, 3500-140C1, 3500-140U1, size 1 -2, 40mm Polymer Liner
Quantity: 240 units
What to do
Check whether you have the product (match the lot numbers, dates or codes in the official notice). Stop using it, and return it or throw it away as the notice says.
Read the official FDA noticeDetails
- Company
- restor3d Inc., Durham, NC
- Date
- Severity
- Class II
the product could cause temporary or reversible health problems. - Where sold
- Nationwide
US Nationwide distribution in the states of Texas, Massachusetts, Oklahoma, Florida, Colorado, North Carolina. - Status
- Ongoing
- Recall number
- Z-0348-2027
- Source
- U.S. Food and Drug Administration