Where's it on?
FDA · Medical devices ·

Tandem Mobi Insulin Pump recall

Tandem Diabetes Care, Inc. recalled Tandem Mobi Insulin Pump on November 11, 2026 because of tandem Diabetes Care, Inc. It was sold nationwide.

Why it was recalled

Tandem Diabetes Care, Inc. ( Tandem ) has identified that certain Tandem Mobi insulin pumps may generate false-positive occlusion alarms when no actual insulin flow obstruction is present. Investigation determined that increased friction within the drive module can reduce available motor power and, under certain conditions, cause motor speed to fall below the occlusion detection threshold. As a result, the pump may incorrectly detect an occlusion condition and suspend insulin delivery. When an…

What's recalled

Tandem Mobi Insulin Pump, Models: 1004004, 1010750, 1012676, 1012719, 1013655, 1013656, 1013501, 1013700, 1014081, 1014801, 1014811, 1014813, 1014936, 1016911, 1016912, 1016913, 1016914, 1018075, 1018077, 1018078

Quantity: 80,938 (77,559 US, 3,379 OUS)

What to do

Check whether you have the product (match the lot numbers, dates or codes in the official notice). Stop using it, and return it or throw it away as the notice says.

Read the official FDA notice

Details

Company
Tandem Diabetes Care, Inc., San Diego, CA
Date
Severity
Class I · most serious
the most serious: the product could cause serious health problems or death.
Where sold
Nationwide
Worldwide - US Nationwide distribution including in the states of CA, FL, NC, OK, UT, WA, IL, PA, MN, VA, NY, IN, HI, OR, GA, MI, KS, CO, WI, SC, null, TX, NE, AZ, OH, NM, NV, MA, SD, IA, TN, ID, MD, RI, CT, NJ, ME, KY, MT, WV, VT, MS, AR, AL, DC, LA, MO, NH, PR, ND, AK, DE, WY and the country of C…
Status
Ongoing
Recall number
Z-0316-2027
Source
U.S. Food and Drug Administration