Tandem Mobi Insulin Pump recall
Tandem Diabetes Care, Inc. recalled Tandem Mobi Insulin Pump on November 11, 2026 because of tandem Diabetes Care, Inc. It was sold nationwide.
Why it was recalled
Tandem Diabetes Care, Inc. ( Tandem ) has identified that certain Tandem Mobi insulin pumps may generate false-positive occlusion alarms when no actual insulin flow obstruction is present. Investigation determined that increased friction within the drive module can reduce available motor power and, under certain conditions, cause motor speed to fall below the occlusion detection threshold. As a result, the pump may incorrectly detect an occlusion condition and suspend insulin delivery. When an…
What's recalled
Tandem Mobi Insulin Pump, Models: 1004004, 1010750, 1012676, 1012719, 1013655, 1013656, 1013501, 1013700, 1014081, 1014801, 1014811, 1014813, 1014936, 1016911, 1016912, 1016913, 1016914, 1018075, 1018077, 1018078
Quantity: 80,938 (77,559 US, 3,379 OUS)
What to do
Check whether you have the product (match the lot numbers, dates or codes in the official notice). Stop using it, and return it or throw it away as the notice says.
Read the official FDA noticeDetails
- Company
- Tandem Diabetes Care, Inc., San Diego, CA
- Date
- Severity
- Class I · most serious
the most serious: the product could cause serious health problems or death. - Where sold
- Nationwide
Worldwide - US Nationwide distribution including in the states of CA, FL, NC, OK, UT, WA, IL, PA, MN, VA, NY, IN, HI, OR, GA, MI, KS, CO, WI, SC, null, TX, NE, AZ, OH, NM, NV, MA, SD, IA, TN, ID, MD, RI, CT, NJ, ME, KY, MT, WV, VT, MS, AR, AL, DC, LA, MO, NH, PR, ND, AK, DE, WY and the country of C… - Status
- Ongoing
- Recall number
- Z-0316-2027
- Source
- U.S. Food and Drug Administration