Senographe Pristina recall
GE Medical Systems, LLC recalled Senographe Pristina on November 11, 2026 because of gE HealthCare has become aware of a nonconformance to certain Senographe Pristina VIA systems. It was sold in USA and OUS..
Why it was recalled
GE HealthCare has become aware of a nonconformance to certain Senographe Pristina VIA systems. Specifically, in cases of 3D+2D manual mode where the acquisition configuration of the anode track/filter combination is different, it can result in the following: the collimator blades might be visible in the image, or the collimator might be open more than expected, and an incorrect collimator filter might be used with respect to the anode track selected This issue could result in a minor additional…
What's recalled
Senographe Pristina
What to do
Check whether you have the product (match the lot numbers, dates or codes in the official notice). Stop using it, and return it or throw it away as the notice says.
Read the official FDA noticeDetails
- Company
- GE Medical Systems, LLC, Waukesha, WI
- Date
- Severity
- Class II
the product could cause temporary or reversible health problems. - Where sold
- USA and OUS.
- Status
- Ongoing
- Recall number
- Z-0327-2027
- Source
- U.S. Food and Drug Administration