Where's it on?
FDA · Medical devices ·

TruSystem 7000 U Operating Table (dV) recall

Baxter Healthcare Corporation recalled TruSystem 7000 U Operating Table (dV) on November 11, 2026 because of the main lift may unexpectedly drop from its set position to its lowest position during use. It was sold nationwide.

Why it was recalled

The main lift may unexpectedly drop from its set position to its lowest position during use. The tabletop may not be able to be raised again after the drop occurs. An uncontrolled drop of the surgical table may result in an abrupt patient displacement during surgery. Serious or critical adverse health consequences may occur, including traumatic injury, airway compromise, and surgical procedure complications, among others.

What's recalled

TruSystem 7000 U Operating Table (dV), Product Code 1723633

Quantity: 62 units

What to do

Check whether you have the product (match the lot numbers, dates or codes in the official notice). Stop using it, and return it or throw it away as the notice says.

Read the official FDA notice

Details

Company
Baxter Healthcare Corporation, Round Lake, IL
Date
Severity
Class II
the product could cause temporary or reversible health problems.
Where sold
Nationwide
US Nationwide distribution in the states of IL, MO, TX.
Status
Ongoing
Recall number
Z-0332-2027
Source
U.S. Food and Drug Administration