Trinias Opera Smart (C16s recall
SHIMADZU CORPORATION MEDICAL SYSTEMS DIVISION recalled Trinias Opera Smart (C16s on November 11, 2026 because of the mounting bolts securing the C-arm drive unit of the X-ray system may not be torqued to specification, potentially compromising structural stability and safe operation. It was sold nationwide.
Why it was recalled
The mounting bolts securing the C-arm drive unit of the X-ray system may not be torqued to specification, potentially compromising structural stability and safe operation.
What's recalled
Trinias Opera Smart (C16s, F12s)/Opera Grand(C12s,C16s, F12s)/Smart OBL Interventional Fluoroscopic X-Ray System containing C-arm support (MH-500, MH-60),
Quantity: 118 unts.
What to do
Check whether you have the product (match the lot numbers, dates or codes in the official notice). Stop using it, and return it or throw it away as the notice says.
Read the official FDA noticeDetails
- Company
- SHIMADZU CORPORATION MEDICAL SYSTEMS DIVISION, Kyoto
- Date
- Severity
- Class II
the product could cause temporary or reversible health problems. - Where sold
- Nationwide
US Nationwide distribution in the states of SC, TX, FL, TN, CA, FL, NY. - Status
- Ongoing
- Recall number
- Z-0347-2027
- Source
- U.S. Food and Drug Administration