Baxter Healthcare Corporation recalled Operating Table TruSystem 7000 U (dV) on November 11, 2026 because of there is an issue with the main lift mechanism. It was sold nationwide.

Why it was recalled

There is an issue with the main lift mechanism. The main lift may unexpectedly drop from its set position to its lowest position during use. In these cases, the tabletop could not be raised again after the drop occurred. The issue has been associated with unexpected wear of the main lift assembly, also known as the telescopic spindle. An uncontrolled drop of the surgical table may result in an abrupt patient displacement during surgery. Serious or critical adverse health consequences may occur,…

What's recalled

Operating Table TruSystem 7000 U (dV), Model/Catalog Number: 1723633

Quantity: 62 units

What to do

Check whether you have the product (match the lot numbers, dates or codes in the official notice). Stop using it, and return it or throw it away as the notice says.

Read the official FDA notice

Details

Company
Baxter Healthcare Corporation, Deerfield, IL
Date
Severity
Class II
the product could cause temporary or reversible health problems.
Where sold
Nationwide
Worldwide - US Nationwide distribution.