Where's it on?
FDA · Medical devices ·

VERIQA; Product Name recall

PTW FREIBURG recalled VERIQA; Product Name on September 23, 2026 because of when a bolus is included in a treatment plan and sent for secondary dose calculation, the software accounts for it but does not display this clearly, which may lead users to incorrectly omit the bolus during treatment and deliver the prescribed dose to the wrong tissue depth. It was sold nationwide.

Why it was recalled

When a bolus is included in a treatment plan and sent for secondary dose calculation, the software accounts for it but does not display this clearly, which may lead users to incorrectly omit the bolus during treatment and deliver the prescribed dose to the wrong tissue depth.

What's recalled

Brand Name: VERIQA; Product Name: RT MonteCarlo 3D v3; Model/Catalog Number: S070031.520; Software Version: 3.1.18;

Quantity: 2

What to do

Check whether you have the product (match the lot numbers, dates or codes in the official notice). Stop using it, and return it or throw it away as the notice says.

Read the official FDA notice

Details

Company
PTW FREIBURG, Freiburg Im Breisgau, N/A
Date
Severity
Class II
the product could cause temporary or reversible health problems.
Where sold
Nationwide
Worldwide - US Nationwide distribution in the states of GA, NY and the countries of Austria, Belgium, Switzerland, Czechia, Germany, Spain, France, Netherlands, Norway, Poland, Portugal, Romania, T¿rkiye, Chile, United Arab Emirates, Saudi Arabia, Mexico, Hong Kong, Brazil, Tajikistan, Ecuador, Ira…
Status
Ongoing
Recall number
Z-3174-2026