Where's it on?
FDA · Medical devices ·

Resmed Astral 100/150 Resmed Astral 100/50 is a ventilator system that provides continuous or intermittent ve… recall

ResMed Ltd. recalled Resmed Astral 100/150 Resmed Astral 100/50 is a ventilator system that provides continuous or intermittent ve… on August 5, 2026 because of due to a Supercapacitor electrolyte leakage on Printed Circuit Board Assembly (PCBA) may damage the charger chip, causing fail-safe state and interruption of ventilation therapy. It was sold nationwide.

Why it was recalled

Due to a Supercapacitor electrolyte leakage on Printed Circuit Board Assembly (PCBA) may damage the charger chip, causing fail-safe state and interruption of ventilation therapy.

What's recalled

Resmed Astral 100/150 Resmed Astral 100/50 is a ventilator system that provides continuous or intermittent ventilatory support for patients weighing more than 11lb (5kg) who require mechanical ventilation.

Quantity: 168,069 devices globally

What to do

Check whether you have the product (match the lot numbers, dates or codes in the official notice). Stop using it, and return it or throw it away as the notice says.

Read the official FDA notice

Details

Company
ResMed Ltd., Bella Vista, N/A
Date
Severity
Class I · most serious
the most serious: the product could cause serious health problems or death.
Where sold
Nationwide
Worldwide - U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY. The countries of Algeria, A…
Status
Ongoing
Recall number
Z-2735-2026
Source
U.S. Food and Drug Administration

More medical devices recalls