Where's it on?
FDA · Medical devices ·

Oncology sets with drip chambers recall

ICU Medical, Inc. recalled Oncology sets with drip chambers on September 23, 2026 because of intravascular administration tubing set drip chambers may have black specks potentially due to the plastic molding process leaving discolored polyvinyl chloride (PVC) material within the drip chamber wall that does not enter the fluid path, which if seen by a healthcare provider may lead to replacement of product before use, which could result in a delay in therapy. It was sold nationwide.

Why it was recalled

Intravascular administration tubing set drip chambers may have black specks potentially due to the plastic molding process leaving discolored polyvinyl chloride (PVC) material within the drip chamber wall that does not enter the fluid path, which if seen by a healthcare provider may lead to replacement of product before use, which could result in a delay in therapy.

What's recalled

Oncology sets with drip chambers

Quantity: 107,800

What to do

Check whether you have the product (match the lot numbers, dates or codes in the official notice). Stop using it, and return it or throw it away as the notice says.

Read the official FDA notice

Details

Company
ICU Medical, Inc., San Clemente, CA
Date
Severity
Class II
the product could cause temporary or reversible health problems.
Where sold
Nationwide
Worldwide - US Nationwide distribution including in the states and territories of ND, TX, CT, KY, CO, AR, NC, NJ, VA, MD, OK, AK, IA, IN, KS, OR, CA, OH, WA, GA, AL, SC, MA, HI, FL, NH, NE, MN, MS, SD, MT, NY, DC, ID, WI, PA, AZ, MO, NV, TN, NM, WV, WY, IL, UT, MI, RI, VT, LA, DE, ME, VI, GU, PR an…
Status
Ongoing
Recall number
Z-3158-2026
Source
U.S. Food and Drug Administration