Contego Medical, Inc recalled Neuroguard IEP 3 in 1 Carotid Stent and Post-Dilation Balloon System with Integrated Embolic Protection (Neur… on August 12, 2026 because of carotid stent and post-dilation balloon system with integrated embolic protection system has an integrated filter, that due to a wire-to-filter-bond issue, may fail to open following sheath retraction; and if this condition were to occur during a procedure, the result may be loss of intended integrated embolic protection. It was sold nationwide.

Why it was recalled

Carotid stent and post-dilation balloon system with integrated embolic protection system has an integrated filter, that due to a wire-to-filter-bond issue, may fail to open following sheath retraction; and if this condition were to occur during a procedure, the result may be loss of intended integrated embolic protection.

What's recalled

Neuroguard IEP 3 in 1 Carotid Stent and Post-Dilation Balloon System with Integrated Embolic Protection (Neuroguard IEP System), REF: NG-0630-140-2, NG-0640-140-2, NG-0730-140-2, NG-0740-140-2, NG-NV-6-30, NG-NV-6-40, NG-NV-7-30, NG-NV-7-40

Quantity: 5069