Fort Defiance Industries (FDI) recall
Fort Defiance Industries, LLC recalled Fort Defiance Industries (FDI) on August 26, 2026 because of the Automated Field Steam Sterilizer's water recovery system 120V Amphenol connector may experience an intermittent loss of ground continuity, which could result in an unreliable fault current return path. It was sold nationwide.
Why it was recalled
The Automated Field Steam Sterilizer's water recovery system 120V Amphenol connector may experience an intermittent loss of ground continuity, which could result in an unreliable fault current return path.
What's recalled
Fort Defiance Industries (FDI), Automated Field Steam Sterilizer REF: P2131
What to do
Check whether you have the product (match the lot numbers, dates or codes in the official notice). Stop using it, and return it or throw it away as the notice says.
Read the official FDA noticeDetails
- Company
- Fort Defiance Industries, LLC, Loudon, TN
- Date
- Severity
- Class II
the product could cause temporary or reversible health problems. - Where sold
- Nationwide
Worldwide - US Nationwide distribution in the states of MD, VA, NJ, TN and the countries of Canada, Lithuania, Spain. - Status
- Ongoing
- Recall number
- Z-2983-2026
- Source
- U.S. Food and Drug Administration
More medical devices recalls
· Olympus Corporation of the AmericasMedline convenience kits containing Healthmark Instrument Channel Brushes
· Medline Industries, LPMidmark IQvitals Zone
· Midmark CorporationAchieva 3.0T; Model Numbers (REF)
· Philips North America LlcAutomated Impella Controller (AIC) with the below product descriptions and corresponding Product Codes. 1
· Abiomed, Inc.INCA (Infant Nasal Cannulae Assy) Neonatal Nasal CPAP, Replacement Set
· CooperSurgical, Inc.