Datex-Ohmeda, Inc. recalled Et Control on September 23, 2026 because of anesthesia Delivery Systems containing certain electronic gas mixer assemblies can automatically transition to Alternate O2 control. It was sold nationwide.

Why it was recalled

Anesthesia Delivery Systems containing certain electronic gas mixer assemblies can automatically transition to Alternate O2 control. In Alternate O2 control, the system delivers 100% O2 to the breathing system enabling continued controlled or manual ventilation and initiates audible and visual alarms. When this issue occurs, anesthetic agent delivery will be interrupted and requires clinician intervention. If this is not provided, the patient may experience decreased anesthetic agent effect.

What's recalled

Et Control, Software option for anesthesia gas machine to achieve and maintain targeted end tidal oxygen and anesthetic agents, Field replaceable unit (FRU) part numbers: 1011-8000-000-S, 5700665

Quantity: 410 units

What to do

Check whether you have the product (match the lot numbers, dates or codes in the official notice). Stop using it, and return it or throw it away as the notice says.