Where's it on?
FDA · Medical devices ·

ERBECRYO 2 Cryosurgical Unit Model/Catalog Number recall

Erbe USA Inc recalled ERBECRYO 2 Cryosurgical Unit Model/Catalog Number on August 19, 2026 because of a strong pull on the high-pressure transfer hose of cryosurgical unit may lead to pneumatic short circuit between he gas inlet and the connector resulting in CO2 leaking into the return path, leading to excessive pressure causing the cryoprobe to rupture. It was sold nationwide.

Why it was recalled

A strong pull on the high-pressure transfer hose of cryosurgical unit may lead to pneumatic short circuit between he gas inlet and the connector resulting in CO2 leaking into the return path, leading to excessive pressure causing the cryoprobe to rupture.

What's recalled

ERBECRYO 2 Cryosurgical Unit Model/Catalog Number: 10402-000

Quantity: 1126 units

What to do

Check whether you have the product (match the lot numbers, dates or codes in the official notice). Stop using it, and return it or throw it away as the notice says.

Read the official FDA notice

Details

Company
Erbe USA Inc, Marietta, GA
Date
Severity
Class II
the product could cause temporary or reversible health problems.
Where sold
Nationwide
US Nationwide distribution including in the states and territories of Alabama, Alaska, Arizona, Arkansas, California, Colorado, Connecticut, Delaware, District of Columbia, Florida, Georgia, Hawaii, Idaho, Illinois, Indiana, Iowa, Kansas, Kentucky, Louisiana, Maine, Maryland, Massachusetts, Michiga…
Status
Ongoing
Recall number
Z-2938-2026
Source
U.S. Food and Drug Administration