Where's it on?
FDA · Medical devices ·

DURACON XL ANGLED TIB INS 11/14MM C-M064-1-300 SELF-PRESSURIZING GLENOID SIZE 8 C-M062-6-800 C-M062-7-500 C-M… recall

Howmedica Osteonics Corp. recalled DURACON XL ANGLED TIB INS 11/14MM C-M064-1-300 SELF-PRESSURIZING GLENOID SIZE 8 C-M062-6-800 C-M062-7-500 C-M… on September 23, 2026 because of affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. It was sold nationwide.

Why it was recalled

Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.

What's recalled

DURACON XL ANGLED TIB INS 11/14MM C-M064-1-300 SELF-PRESSURIZING GLENOID SIZE 8 C-M062-6-800 C-M062-7-500 C-M063-2-600 SELF-PRESSURIZING GLENOID SIZE 9 C-M063-5-000 SHOULDER - GLENOID #11 5361-6111 SHOULDER - GLENOID #5 5361-6105 SHOULDER - GLENOID #7 5361-6107 SHOULDER - GLENOID #9 5361-6109 SHOULDER - KEELED GLENOID #11 5361-6211 SHOULDER - KEELED GLENOID #5 5361-6205 SHOULDER - KEELED GLENOID #7 5361-6207 SHOULDER - KEELED GLENOID #9 5361-6209 Orthopaedic implant components manufactured from UHMWPE (polyethylene)

Quantity: 12,156 US

What to do

Check whether you have the product (match the lot numbers, dates or codes in the official notice). Stop using it, and return it or throw it away as the notice says.

Read the official FDA notice

Details

Company
Howmedica Osteonics Corp., Mahwah, NJ
Date
Severity
Class II
the product could cause temporary or reversible health problems.
Where sold
Nationwide
Worldwide distribution - US Nationwide and the countries of Canada, LATAM, Europe, Asia Pacific, and South Africa.
Status
Ongoing
Recall number
Z-3200-2026
Source
U.S. Food and Drug Administration

More medical devices recalls