COAPTIUM CONNECT Product Name recall
TISSIUM SA recalled COAPTIUM CONNECT Product Name on August 12, 2026 because of currently labeled shelf life of 36 months is not substantiated by real-time data. It was sold nationwide.
Why it was recalled
Currently labeled shelf life of 36 months is not substantiated by real-time data. Devices will be relabeled to reflect revised shelf life of 26 months. The use of a medical device beyond its validated shelf life may result in perioperative breakage of the device rendering device unusable; worst-case outcome would be revision surgery.
What's recalled
Brand Name: COAPTIUM CONNECT Product Name: COAPTIUM CONNECT Model/Catalog Number: NCOCO2010 Software Version: N/A Product Description: In situ polymerizing peripheral nerve repair device Component: No
What to do
Check whether you have the product (match the lot numbers, dates or codes in the official notice). Stop using it, and return it or throw it away as the notice says.
Read the official FDA noticeDetails
- Company
- TISSIUM SA, Paris 12e Arrondissement
- Date
- Severity
- Class II
the product could cause temporary or reversible health problems. - Where sold
- Nationwide
US Nationwide distribution in the states of CA, FL, IL, NJ, OH, PA. - Status
- Ongoing
- Recall number
- Z-2903-2026
- Source
- U.S. Food and Drug Administration
More medical devices recalls
· Olympus Corporation of the AmericasMedline convenience kits containing Healthmark Instrument Channel Brushes
· Medline Industries, LPMidmark IQvitals Zone
· Midmark CorporationAchieva 3.0T; Model Numbers (REF)
· Philips North America LlcAutomated Impella Controller (AIC) with the below product descriptions and corresponding Product Codes. 1
· Abiomed, Inc.INCA (Infant Nasal Cannulae Assy) Neonatal Nasal CPAP, Replacement Set
· CooperSurgical, Inc.