Boston Scientific Corporation StoneSmart Connect Console Contents recall
Boston Scientific Corporation recalled Boston Scientific Corporation StoneSmart Connect Console Contents on September 23, 2026 because of a subset of power adapters used with the connect console may exhibit higher-than-expected resistance in the electrical grounding path. It was sold nationwide.
Why it was recalled
A subset of power adapters used with the connect console may exhibit higher-than-expected resistance in the electrical grounding path. As a result, the power adaptor may fail electrical safety testing during installation.
What's recalled
Boston Scientific Corporation StoneSmart Connect Console Contents: (1) Power Cord, (1) Power Adaptor, (1)DVI Cable, (1) DVI to HDMI Cable, (1) DisplayPort Cable REF: M0067933000, M006793300H0, M006793300R0, M006793300D0, M006793300Z0, M0067930000, M006793000R0
Quantity: 982 units
What to do
Check whether you have the product (match the lot numbers, dates or codes in the official notice). Stop using it, and return it or throw it away as the notice says.
Read the official FDA noticeDetails
- Company
- Boston Scientific Corporation, Marlborough, MA
- Date
- Severity
- Class II
the product could cause temporary or reversible health problems. - Where sold
- Nationwide
Worldwide - US Nationwide distribution including in the states of AZ, FL, CA, OR, IL, NC, NE, UT, CO, OH, AR, NM, AL, WI, TN, IN, GA, OH, WI, KY, CT, KS, SC, IL, NE, MO, MS, DE, TX, NY, WV, MA, MT, WA, WY, NV, HI, ID, MO, ME, WV, NE, OK, MN, LA, WV, NV, PA, NH, NJ, MI, MO, AK, VA, MS, WY, ME and th… - Status
- Ongoing
- Recall number
- Z-3188-2026
- Source
- U.S. Food and Drug Administration
More medical devices recalls
· Olympus Corporation of the AmericasMedline convenience kits containing Healthmark Instrument Channel Brushes
· Medline Industries, LPMidmark IQvitals Zone
· Midmark CorporationAchieva 3.0T; Model Numbers (REF)
· Philips North America LlcAutomated Impella Controller (AIC) with the below product descriptions and corresponding Product Codes. 1
· Abiomed, Inc.INCA (Infant Nasal Cannulae Assy) Neonatal Nasal CPAP, Replacement Set
· CooperSurgical, Inc.