Beckman Coulter DxC 500i Clinical Analyzer Modules using software version 1.5 or lower recall
Beckman Coulter Inc. recalled Beckman Coulter DxC 500i Clinical Analyzer Modules using software version 1.5 or lower on August 26, 2026 because of a software defect in which the Parameter Package (PP) import logic overwrites all predefined calculated test rules, including those previously modified by the user, upon import of a new Parameter Package version. It was sold nationwide.
Why it was recalled
a software defect in which the Parameter Package (PP) import logic overwrites all predefined calculated test rules, including those previously modified by the user, upon import of a new Parameter Package version. As a result, usercustomized calculated test rules revert silently to their default expressions, potentially generating erroneous patient test results without an explicit error notification. The issue affects instruments running software version 1.5 or lower and only occurs when a new P…
What's recalled
Beckman Coulter DxC 500i Clinical Analyzer Modules using software version 1.5 or lower: 1) DxC 500 AU Module w/ISE, DxC 500i, Part Number C63522; 2) DxC 500 AU Module, DxC 500i, Part Number C63521; 3) Access 2 Module, DxC 500i, Part Number C13252; DxC 500 AU Clinical Chemistry Analyzers
What to do
Check whether you have the product (match the lot numbers, dates or codes in the official notice). Stop using it, and return it or throw it away as the notice says.
Read the official FDA noticeDetails
- Company
- Beckman Coulter Inc., Brea, CA
- Date
- Severity
- Class II
the product could cause temporary or reversible health problems. - Where sold
- Nationwide
Worldwide distribution - US Nationwide. - Status
- Ongoing
- Recall number
- Z-2973-2026
- Source
- U.S. Food and Drug Administration
More medical devices recalls
· Olympus Corporation of the AmericasMedline convenience kits containing Healthmark Instrument Channel Brushes
· Medline Industries, LPMidmark IQvitals Zone
· Midmark CorporationAchieva 3.0T; Model Numbers (REF)
· Philips North America LlcAutomated Impella Controller (AIC) with the below product descriptions and corresponding Product Codes. 1
· Abiomed, Inc.INCA (Infant Nasal Cannulae Assy) Neonatal Nasal CPAP, Replacement Set
· CooperSurgical, Inc.